Load cell damage diagnosis should end with evidence for repair, replacement or further testing. A shifted reading alone does not identify a damaged sensing element. The same symptom can arise from a trapped cable, a strained mount, moisture in a junction box or a changed instrument setting. If replacement is justified, the next question is whether the proposed sensor will fit the existing assembly and whether the cause of damage has been removed.

Load cell damage diagnosis: distinguish a suspect from a confirmed fault
Begin with the findings from the system troubleshooting sequence. Record what has already been ruled out and what remains uncertain. A channel that behaves differently deserves investigation, but the channel includes mounting, sensor, cable and connections. Avoid reducing that evidence to "cell failed" before the fault location is established.
| Finding | Reasonable conclusion | Required follow-up |
|---|---|---|
| Visible deformation or damaged loading parts | The assembly needs assessment before further use. | Inspect the sensor and surrounding structure under an approved safe work method. |
| Abnormal isolated bridge resistance | An electrical fault is possible in the sensor or attached lead. | Verify pin identification, connections, test conditions and model-specific data. |
| Normal bridge resistance but persistent measurement error | The resistance check has not established measurement performance. | Evaluate zero, load response, mounting and any appropriate specialist tests. |
| Fault appears only in the installed channel | The installation or signal path remains a candidate cause. | Compare controlled tests without bypassing the very section under suspicion. |
Stop using a visibly damaged support or sensor until competent personnel have assessed the installation. A load cell may also be part of the mechanical support path, so removal is not just an electrical repair. Unload, isolate and support the structure using an approved method before releasing mounting fasteners.
Use resistance data as a diagnostic reference, not a certificate
The METRAVON MV-LC0030 lists input resistance of 400 +/-20 ohm and output resistance of 380 +/-5 ohm. The MV-LC0190 lists 350 +/-5 ohm for both. Input and output resistance are therefore not universally equal, and comparing an unidentified bridge with an assumed 350-ohm value can produce a false diagnosis.
These published values describe the relevant models. They are not a universal in-service pass/fail test. Account for the identified terminals, attached cable, temperature, instrument accuracy and any remaining parallel circuit. Follow the manufacturer-approved method with power isolated and connected electronics protected.
A normal resistance value does not prove correct sensitivity, acceptable zero balance or insulation under operating conditions. Likewise, an unusual reading at the indicator terminals may include a wiring fault elsewhere. Document where the measurement was made and compare it with the supplied unit data and earlier results when available.
Do not improvise a high-voltage insulation test from a generic internet procedure. Obtain the permitted voltage, connections and limits for the specific sensor, with the instrument disconnected as required. If those instructions or suitable equipment are unavailable, use qualified test support rather than risk damaging a serviceable unit.
Investigate why the damage happened
An overload record should include the loading event and direction, not only the largest displayed weight. A short impact or side load may not appear as a recorded peak at a filtered indicator. Inspect stops, alignment, loading pieces and the process that produced the event before installing another sensor.
The MV-LC0030's published 200% F.S. safe-load reference is not permission to operate above rated capacity. It does not establish an allowed side load, impact duration or repeated-overload duty. Staying below that number on the display cannot, by itself, rule out damage.
Where moisture is involved, check the entire path: cable entry, splice, connector and junction box. Where a mechanical obstruction is found, correct it before recalibration. Replacing the sensing element while retaining the same leak or misaligned loading piece leaves the new unit exposed to the same mechanism.

Turn the diagnosis into a replacement specification
Identify the original sensor, capacity and serial number, and obtain the mounting drawing. Compare load direction, height, threads or holes, bearing surfaces, cable route and available removal space. Include the mounting parts in the enquiry if they are worn or damaged. An equal tonnage and similar photograph are not a compatibility check.
Then compare excitation, rated sensitivity, bridge resistance and the instrument input. In a multi-cell installation, assess the replacement alongside the remaining sensors and the balancing method. A connector with the same shape can have a different pin assignment. Neither wire colour nor plug fit should be used as the sole wiring reference.
For an MV-LC0190 replacement, the 1 kg version has a different nominal sensitivity from the 5-50 kg versions. Preserve the capacity-specific electrical data with the order. For an MV-LC0030, confirm the dimensional group and loading interfaces. For a vehicle scale, the weighbridge replacement guide covers the wider deck and support checks.
Send METRAVON the failed-unit identification, drawings, test results, suspected damage event and connected instrument through the replacement enquiry page. State whether you need a sensor alone or the associated mounting parts, and ask for any differences from the original configuration to be listed in the offer.
Verify the repaired installation before release
Record the replacement serial and its position. Check the restored channel, then verify the complete scale using the agreed reference loads and operating conditions. Include repeatability, return to zero, relevant loading positions and the final output used by the controller. Zeroing alone does not verify the span or how the supports share load.
Confirm that temporary supports and test overrides have been removed under the site's approved procedure. Retain the failed component and its records where further investigation or a warranty assessment is needed. Close the repair with evidence that the original symptom no longer occurs and that the installation meets its acceptance requirement.
Questions before ordering the replacement
Q1. Should all sensors be replaced when one is damaged?
Not automatically. Assess the remaining units, the event that affected the system and the compatibility of a single replacement. A common overload or environmental exposure may justify wider inspection. Replace a complete set when the evidence or documented system design requires it, not merely because individual matching has not been checked.
Q2. Can a damaged attached cable be repaired instead of replacing the cell?
Possibly, but the repair must be permitted for that construction and service. It can affect sealing, resistance, shielding and calibration, and some cable entries are not field-serviceable. Obtain an approved repair method and verification requirements before cutting or splicing the lead.
Summary
Replacement approval needs a credible fault finding, a matching mechanical and electrical specification, and a plan to verify the installed result. Correct the damage mechanism at the same time. A new sensor is a replacement part, not evidence that the original fault has been resolved.